Technical requirements for housings for surgical lamps
1. The surface should be flat and
free of defects such as bumps. The light-transparent holes and vent holes on
the round end face should be evenly distributed and unblocked.
2. The appearance level of the
electroplated parts of the surgery light price shall comply
with the requirements of the light level 2 of YY0076.
3. The appearance of the aluminum
oxide film of the shadowless lamp shall comply with the Class 2 requirements
specified in WS2-100.
4. The appearance of the
shadowless paint parts shall comply with the Class 2 requirements specified in
ZBC30003.1.
5. The climatic environment test
and mechanical environmental test of the medical lights
factory shall comply with the climatic environment test and
mechanical environmental test group I of WS2-283 (in the case of working high
temperature, working heat and humidity, the temperature rise test is not
performed)