The wrong way of using surgical light and how to maintain it
As a medical device which is widely used in
medical units today, the surgical lamp LED
has the advantages of low maintenance cost, simple construction, high safety
and long service life, so it has gradually won the recognition and favor of
more and more people in the industry. However, it is still very difficult to
avoid the damage caused by the wrong operation. Therefore, it is very important
to know in advance the damage caused by the wrong operation of the halogen shadowless lamp
and to do the inspection work. Below, it is up to Mingtai company to give you a
simple science, I hope everyone will be interested.
Operating light wrong operation:
1.
The height of the lamp head
does not fit the proper position; the lamp head is at least 2 meters above the
ground.
2.
The base is not properly placed,
there is no guarantee that the respective functions do not interfere with each
other, it is not convenient for the lamp head to rotate, or the rotation is
easy to collide with other objects.
3.
Use corrosive liquids to clean
surgical shadowless lamps, which will damage the surface of the lamp body.
4.
Other objects are randomly
added on the balance arm of the surgical lamp, which will affect the balance of
the surgical lamp arm.
5.
Frequently wipe the mirror
surface, which is easy to wear and affect the life of the mirror surface.
6.
Frequent switching of the
surgical light can have an adverse effect on the surgical light supply module
and the bulb body.
7.
After the shadowless lamp is
used, it is not checked and maintained.
Surgery light inspection maintenance
usually includes the following items:
1.
Mechanical installation check:
check for mechanical installation errors, whether there are fewer screws or
screws is not tight, whether the decorative covers have been covered.
2.
Circuit check: First check if
the system has short circuit or open circuit. If everything is normal, you can
conduct power test. Check whether the external power supply is stable, and whether
the output voltage of the transformer is too high or too low.
3.
Balance arm adjustment: Check
whether the balance arm and the lamp head match, whether it is necessary to
adjust the force of the balance arm, whether to adjust the angle of the balance
arm, usually the screw which is used to adjust the lamp arm is placed at the
two ends of the lamp arm, but different manufacturers are slightly different, observe
carefully.
4.
Joint flexibility adjustment:
mainly to adjust the joint's damping screw. The standard specifies that the
tightness degree of the damping adjustment is that the force is close to 20N or
5Nm when pushing or rotating the joint at any position.