LED shadowless lamp is a medical lighting tool used by medical personnel to observe the
incision of the surgical field and different depths of tissue in the body
cavity. Shadowless lights often have a variety of sudden failures during use. Mingtai
Company summarized the common faults of LED
operating light for everyone.
1. The main light is off and the auxiliary
light is on. ZF series operation light has automatic switching function in the
circuit control. When the main lamp is damaged, the auxiliary lamp will light
up to ensure normal operation. After surgery, the main lamp bulb should be
replaced immediately.
2. Light is not on. Open the top cover,
check the transformer and power supply voltage to see whether thay are normal,
if both are no problem, please find a professional to repair.
3. Transformer damage. There are generally
two reasons for transformer damage: Supply voltage and excessive current caused
by short circuit. The latter should ask a professional to repair.
4. The bulb is often damaged. Check whether
the bulb is configured according to the rated power specified in the
instruction manual and whether the power supply voltage is normal.
5. Disinfection handle deformation. Surgery light
handle can be autoclaved, but the handle can not press heavy objects during
disinfection, otherwise it will cause the handle deformation.
6. Move operating light angle, the light is
off. This is mainly because the sensors at both ends of the arm will cause poor
contact after a period of use. In this case, the professional should be asked
to maintain it once.
7. Lamp head drift. After using for a
period of time, due to the heavy weight of the operating lamp head, a large
friction force is required for positioning, lamp head drift is generated, the
upper positioning screw can be tightened to solve the problem.
8. Surgery
shadowless light dims. Reflective glass for cold light shadowless
lamps using coating technology. Generally domestic coating technology has only
two years of life, after two years, the coating layer will cause problems, the
reflection will become dark and foaming. In this case, you need to change the
reflective bowl.