How to measure the shadowlessness of surgery light
Use 50mmxX75mm black cylinder, place light
source 1 meter above it, place a photometer at the bottom of the black cylinder,
the measured light/shadow ratio must not be less than 10%, the actual light
source used is surgery light with
diameter of 50cm to 100cm, only some major performance indicators meet the
standards then it can be used. Here are some important performance parameters
listed, we can refer to the following standards when buying.
First of all, the most important point is
to have sufficient illumination. Now surgery
lamp illumination can reach 180,000 Lux, near to the sun brightness
in summer. The actual use of illumination generally is between 40,000 – 100,000
Lux, too bright will affect eyesight. Glare on surgical instruments should be
avoided while providing adequate illumination, glare affects vision and vision,
easy to make the eyes fatigue, adverse surgeon surgery. Surgery lamp
illumination and operating room illumination should not have a big difference.
Some standards require that the illumination of the overall illumination is
one-tenth of the local illumination. The operating room temperature increase
should less than 10 Celsius. The main way is to make the visible light
generated by the light source through the filter which can filter out 70%
infrared, most of the remaining infrared rays are separated and filtered by
glass cover. After the above treatment the beam is called cold light.
Secondly, the shadowless lamp has a high
degree of shadowlessness, and the shadowlessness is an important feature and
performance index of the surgical shadowless lamp. Operating light install a
few or a dozen ring evenly arranged light source in 50-100 cm diameter lamp
head, special bulbs and reflectors are used in the light source, the beam which
from different points meets at one point through filter reflection, to provides
balanced, uniform and shadowless bright beam.